GE CRx25P Brochure Pagina 51

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 114
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 50
99
In this situation the unsharp images of each of the two edges of the defect may overlap,
as shown in example C. The result is that image C not only becomes unsharp, but also
suffers a reduction in contrast compared to image A, made with a point source and
image B made with a relatively small source.
Inherent unsharpness
Not only the silver halide crystals directly
exposed to X-radiation are formed into
grains of silver, but also (albeit to a lesser
degree) the surrounding volume of emul-
sion. This cross-sectional area represents
the “inherent unsharpness” or “film uns-
harpness” U
f
.
So, even in the absence of geometric uns-
harpness, if the radiation energy is high
enough, film unsharpness can occur: the so
called “inherent unsharpness”. If a steel test
plate with a sharp thickness transition is
radiographed with high energy X-rays,
there will be a gradual transition of film
density across the image of the “step” from
A to B.
Without inherent unsharpness, the film
would show an absolutely sharp transition
between the two densities, as shown in figu-
re 3a-11. In practice, the density change
across the image is as shown in figures 3b,
3c and 3d-11.
The width of this transitional area (U
f
),
expressed in mm, is a measure of film
unsharpness.
98
Consequently, U
g
can be reduced to any required value by increasing the source-to-film
distance. However, in view of the inverse square law this distance cannot be increased
without limitation, as extremely long exposure-times would result. The formula also
indicates that geometric unsharpness assumes more and more importance as the distan-
ce between defect and film increases.
A special case arises, however, when one uses a micro focus X-ray tube with a focal spot
size in the range 10-50
m. With such a small focus size, the image can be deliberately
magnified (see section 17.1) by using a short source-to-specimen distance, and a large
specimen-to-film distance, and still retain an acceptably small value of U
g
. The advan-
tage of this technique, called the “projective magnification method”, is that the grainin-
ess always present in a photographic image is less of a disturbing factor in the discerni-
bility of very small defects.
Figure 2-11 shows the effect of geometric unsharpness on the image of a defect smaller
than the focus size.
film film film
Fig. 2-11. Geometric unsharpness: effect on the image of a small defect.
A. Point focus size - s -: no geometric unsharpness - defect image sharp
B. Small focus size – s -: geometric unsharpness U
g
– defect image blurred
C. Increased focus size – s -: still larger Ug - defect image blurred and loss of contrast - C
o
is less than in A and B
C
o
= contrast
Fig. 3-11. Inherent (film) unsharpness for X and
Gamma-radiation.
For clarity, the density curves are magnified along the X-axis.
(a) density distribution across image of sharp edge,
assuming U
f
= 0
(b) (c) and (d) density distribution due to film unsharpness
(b) theoretical; (c) with grain; (d) smoothed.
film
test plate
film density
Vedere la pagina 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 113 114

Commenti su questo manuale

Nessun commento