April 2004
3-22
Installing DMP-1 Dampening Material with CFG-V
Series Fixtures (cont.)
7. With a marker, trace around the transducer footprint, as shown
in Figure 3-24 below.
Figure 3-24: DMP-1 Material with
Transducer Footprint Traced
8. Remove the fixture and transducer. Then use a utility knife to
cut out the area under the transducer footprint, and peel the cut
material off the pipe, as shown in Figure 3-25 below.
Figure 3-25: DMP-1 Material with
Transducer Footprint Cut Out
9. To remove any residual adhesive, clean the cut-out area with a
rag or a piece of sandpaper.
10. Reinstall the fixture and transducers on the pipe.
Vedere la pagina 53